UxOATN3024
|
UxOATN 24W Series is an embedded open-frame power supply for medical, home appliance, and ICT applications. Featuring a high-insulation design compliant with 2x MOPP and high efficiency meeting Energy Efficiency Level VI, it ensures stable operation. It includes short-circuit, overvoltage, overcurrent, and overtemperature protection, and its compact 2x3 inch board is ideal for built-in equipment design. Products Series Name:UxOATN3024 |
|
Products Details
| Size | L:76.2 W:50.8 H:31.1/mm |
|---|---|
| AC Input Rating | 90V~264V |
| DC Output Rating |
|
| Safety |
|
| Application Examples by Use Case | Introduction to Products by Application (PDF) |
If you are looking for a Products with specifications other than those listed above, please contact our sales team.
Contact Form
The UxOATN 24W Series is an open-frame AC/DC power supply (24W) compatible with medical equipment, home appliances, and ICT equipment. Depending on the safety standard classification, three types are available: M (60601-1) for medical equipment, H (61558-1/60335-1) for home appliances, and N (62368-1/60065-1/60950-1) for ICT equipment. For medical applications, it supports 2xMOPP (2 Means of Patient Protection) and can obtain CB and UL certification.
Featuring an open-frame design with a 2"x3" board size (approx. 50.8x76.2mm), the series offers four models: 5V/12V/15V/24V. In addition to compliance with PSE Appendix Table 8 (100V standard) and dust requirements, it supports over-temperature protection (optional) and inrush current control, ensuring long-term stable operation in a wide range of installation environments such as medical facilities, factories, and homes.
| UNIFIVE is chosen because: As a power supply manufacturer with over 35 years of development experience, we provide comprehensive support from specification planning for mass production, certification acquisition, mass production launch, to post-production follow-up. Based on standard products, we offer flexible supply options according to production volume, from small-lot prototyping and evaluation to small- and medium-scale production and large-volume manufacturing. We also support semi-custom solutions and assist with power supply selection from the specification planning stage for mass production adoption. |
Key Features
1) Supports three applications: medical, home appliance, and ICT; 2xMOPP medical-grade insulation
By specifying the safety standard classification in the model number, you can choose from three types: M (60601-1) for medical equipment, H (61558-1/60335-1) for home appliances, and N (62368-1/60065-1/60950-1) for ICT equipment. The medical-grade M series supports 2xMOPP (2 Means of Patient Protection) and can obtain CB and UL certification. Select the H series for IEC 60335-compliant equipment and home appliances, and the N series for ICT and AV equipment. Please contact our sales department for detailed certification status of each model.
2) 2"x3" board size (approx. 50.8x76.2mm); four models from 5V to 24V
The 2"x3" board size (approx. 50.8x76.2mm) open-frame design is suitable for internal installation in equipment. Output voltages are available in four models: 5.0V/12.0V/15.0V/24.0V, supporting a wide range of applications from low-voltage logic systems to 15V analog systems and 24V actuators and sensors. Equipped with input connector CN1 (equivalent to JST B3P-VH) and output connector CN2 (equivalent to JST B4B-XH, 4-pin).
3) Over-temperature protection (optional) and inrush current control
In addition to short-circuit, overvoltage, and overcurrent protection, over-temperature protection (OTP) is available as an option. By detecting abnormal temperature rise and activating automatic protection, it enhances the safety of medical equipment and devices operating continuously for long periods. It also supports inrush current control, reducing the impact of transient current on peripheral circuits at power-on.
4) Comprehensive EMC compliance covering medical, home appliance, and ICT standards
For emission standards, it complies with FCC Part15-B, FCC Part18-B, CE (EN/CISPR55032-B, EN/CISPR55011-B, EN/CISPR55014-B), BS EN 55032, BS EN55011, BS EN55014-1, and VCCI-B. Immunity standards include EN55035, BS EN 55035 (ICT), EN60601-1-2, BS EN60601-1-2 (medical), EN55014-2, and BS EN55014-2 (home appliances), covering all three application categories. Comprehensive compliance includes EN61000-4-2 to 4-6, 4-8, and 4-11.
5) PSE Appendix Table 8 compliant, 100V standard compliant, dust requirement compliant, 5-year power supply lifespan option
Complies with PSE Appendix Table 8 (100V standard), making it suitable as a built-in power supply under Japan's Electrical Appliance and Material Safety Law. Compliance with dust requirements enables use in dusty installation environments. Designed for energy efficiency in compliance with Energy Efficiency Level VI (ErP/DoE), with an operating temperature range of 0 to 50°C. Additionally, a 5-year power supply lifespan warranty option is available to support adoption in medical and industrial equipment requiring long-term operation.
Main Specifications
| Series Name | UxOATN 24W Series |
| Products Category | Open-frame AC/DC power supply (for medical equipment, home appliances, and ICT equipment; built-in type) |
| Rated Output | 24W |
| Key Features | Supports three applications (medical, home appliance, ICT); 2xMOPP support (M series); 2"x3" board; four models from 5V to 24V; optional over-temperature protection; inrush current control; PSE Appendix Table 8 and dust requirement compliant; 5-year lifespan option |
| Safety Standard Classification and Applicable Standards | M: 60601-1 (medical equipment, 2xMOPP, CB and UL certification available) / H: 61558-1, 60335-1 (home appliances) / N: 62368-1, 60065-1, 60950-1 (ICT equipment), PSE Appendix Table 8 100V standard compliant (contact sales for details per model) |
| Operating Temperature Range | 0 to 50°C (natural air cooling, at sea level) |
| Energy Efficiency | Energy Efficiency Level VI (ErP/DoE), compliant with EU Commission Regulation 2019/1782, DOE 10 CFR Part 429 and 430 |
| Emissions | FCC Part15-B, Part18-B / CE EN(CISPR)55032-B, 55011-B, 55014-B / BS EN 55032, 55011, 55014-1 / VCCI-B |
| Immunity | EN55035, BS EN 55035 (ICT) / EN60601-1-2, BS EN60601-1-2 (medical) / EN55014-2, BS EN55014-2 (home appliances) (including EN61000-4-2 to 4-6, 4-8, 4-11) |
| Protection Circuits | Short-circuit protection (SCP) / Overvoltage protection (OVP) / Overcurrent protection (OCP) / Over-temperature protection (OTP, optional) |
| Operating Environment | Input: AC100-240V (range: AC90-264V), 50/60Hz (range: 47-63Hz) / Storage temperature: -20 to 65°C (natural air cooling, at sea level) / Operating and storage humidity: 5-95%RH (non-condensing) |
| Reference: Output Voltage, Current, and Accuracy | 5.0V 3.0A (ripple 150mV, noise 150mV) / 12.0V 2.0A (ripple 240mV, noise 240mV) / 15.0V 1.6A (ripple 240mV, noise 240mV) / 24.0V 1.0A (ripple 240mV, noise 240mV) (output accuracy ±5%, output connector CN2) |
| Connector Specifications | CN1 (input: JST B3P-VH equivalent, AC L/N), CN2 (output: JST B4B-XH equivalent, 4-pin, VO+: pins 1-2 / VO-: pins 3-4) |
| Dimensions | 76.2x50.8mm (3"x2"), board thickness 1.6mm, maximum component height 26.5mm (tolerance ±0.5mm) |
| Others | LPS certified, supports inrush current control, 5-year lifespan option available, RoHS 2 compliant |
Semi-Custom Support
Please feel free to contact us for specification confirmation, quotations, or sample requests. Examples of semi-custom solutions can be found on the Customization Case Studies page.