UMOAT3036 P Model with peak-load capability
|
This is a 36W open-frame switching power supply for medical equipment. It complies with IEC 60601-1 and supports 2xMOPP for patient protection and 2xMOOP for operator protection. This compact onboard power supply is ideal for integration into medical equipment. As a Model with peak-load capability, it is ideal for medical equipment that drives DC motors and for dental equipment. Products Series Name:UMOAT3036 P Model with peak-load capability |
|
Products Details
| Size | L:103.0 W:47.0 H:31.6/mm |
|---|---|
| AC Input Rating | 90V~264V |
| DC Output Rating |
|
| Safety |
|
| Application Examples by Use Case | Introduction to Products by Application (PDF) |
If you are looking for a Products with specifications other than those listed above, please contact our sales team.
Contact Form
The UxOAT 36W P Series (UxOAT3036P) is an open-frame power supply (PCB power supply) for medical devices suitable for integration into equipment. Designed for medical, home appliance, and I.C.T. applications, its space-saving 1.9" x 4.1" PCB design facilitates the selection of an internal power supply based on mounting requirements.
As an AC-DC switching power supply, it features short-circuit, overvoltage, and overcurrent protection circuits and supports design and evaluation based on emissions requirements (FCC/EN(CISPR)/BS EN) and immunity requirements (EN55035, EN61000-4-x, etc.). In addition, support for peak loads (1.2 seconds) helps you define power supply requirements while accounting for load fluctuations.
| Why Choose UNIFIVE: As a power supply manufacturer with more than 35 years of development experience, we provide consistent support from specification review and certification to production launch and post-production support, with mass production in mind. Based on our standard products, we offer flexible supply to meet a range of volume requirements, from small-lot prototyping and evaluation to small- and medium-scale production and high-volume production.We also offer semi-custom solutions and support power supply selection from the specification review stage through mass-production adoption. |
Key Features
1) Open-Frame Power Supply for Medical Devices (PCB Power Supply)
This open-frame power supply for medical devices is designed for integration as an internal power supply. Its space-saving 1.9" x 4.1" PCB design makes it suitable for use even where mounting space inside the enclosure is limited.
2) Technical Standards and Protection Circuits
The applicable safety standards vary by model category (M/H/N). Please check the details for each model. Protection circuits include short-circuit protection, overvoltage protection, and overcurrent protection.
3) Supports Peak Loads (1.2 Seconds)
The peak-load function supports up to 170% of the Rated Output during peaks (peak duration: 1.2 seconds, duty: 0.2). This facilitates requirements definition when designing and evaluating internal power supplies for anticipated load fluctuations.
4) Efficiency
Supports Energy Efficiency Level VI (ErP/DoE), enabling selection based on the requirements of Commission Regulation (EU) 2019/1782 and DOE 10 CFR Parts 429 and 430.
5) EMC (Emissions)
Supports FCC Part 15/18 Class B, EN(CISPR)55032/55011/55014, BS EN 55032/55011/55014-1, and other standards. Please refer to "Standards and Compliance" for details.
6) Immunity
Based on the requirements of EN55035/BS EN 55035, EN60601-1-2, EN55014-2, and other standards, immunity tests under EN61000-4-2/-3/-4/-5/-6/-8/-11 are specified. Please refer to "Standards and Compliance" for details.
Effective Use Cases
- Integrating a 36W-class open-frame power supply for medical devices as an internal power supply
- Selecting a power supply based on a 1.9" x 4.1" PCB where mounting space is limited
- Defining power supply requirements while accounting for peak loads (1.2 seconds) caused by load fluctuations during startup or operation
- Reducing the risk of operational issues with a power supply featuring short-circuit, overvoltage, and overcurrent protection
- Selecting a power supply in accordance with emissions requirements (FCC/EN(CISPR)/BS EN) for EMC design
- Selecting a power supply based on immunity requirements (EN61000-4-x), taking disturbances such as ESD, EFT, and surges into account
Applications and Use Examples
Suitable for a wide range of applications as an internal power supply, including medical devices, home appliances, I.C.T. equipment, and AV equipment.
- Equipment requiring an internal power supply
- Medical devices
- Home appliances and consumer electronics
- Information and communications equipment/peripherals, etc.
* The above are examples and do not limit the possible applications.
Standards and Compliance
| Item | Details |
|---|---|
| Technical Standards | M: 60601-1/H: 61558-1 and 60335-1/N: 62368-1, 60065-1, and 60950-1; compliant with 100V standards (varies by model) |
| Peak Load | Up to 170% (peak duration: 1.2 seconds, duty: 0.2) |
| Efficiency | Energy Efficiency Level VI (ErP/DoE)/Commission Regulation (EU) 2019/1782/DOE 10 CFR Parts 429 and 430 |
| EMC (Emissions) | FCC Part 15/18 Class B, EN(CISPR)55032/55011/55014, BS EN 55032/55011/55014-1 |
| Immunity | EN55035/BS EN 55035, EN60601-1-2, EN55014-2, EN61000-4-2/-3/-4/-5/-6/-8/-11 |
| Protection Circuits | Short-circuit protection/overvoltage protection/overcurrent protection |
| Operating Environment | Input voltage: AC100-240V (allowable range: AC90-264V), input frequency: 50/60Hz (allowable range: 47-63Hz)/operating temperature: 0-50°C, storage temperature: -20-65°C, operating/storage humidity: 5-95%RH (non-condensing) |
| Reference: Representative Model | 12V 3.0A (5.1A peak)/15V 2.4A (4.1A peak)/19V 1.9A (3.2A peak)/24V 1.5A (2.5A peak) |
| Other | 2 x MOPP, LPS, RoHS 2 compliant, 5-year service life (optional) |
Semi-Custom Solutions
Please feel free to contact us regarding specification confirmation, quotations, or samples. Examples of semi-custom solutions are available on the Customization Examples page.