UMOYT3036
|
The UMOYT 36W Series is a 36W-class open-frame embedded power supply designed primarily for medical equipment applications and compliant with IEC 60601-1. Its insulation design, compliant with 2 x MOPP, ensures a high level of safety even in medical devices intended for patient contact, enabling power supply designs that meet stringent medical safety requirements. In addition, its high-efficiency design complies with Energy Efficiency Level VI (ErP/DoE), combining energy-saving performance with a stable power supply. It features short-circuit, overvoltage, overcurrent, and overtemperature protection and complies with various EMC/immunity standards, including IEC 60601-1-2. The 2 x 4-inch board provides ample capacity for 36W-class output while offering design flexibility for integration into medical equipment. The UMOYT 36W Series is a power supply solution that supports the design of medical devices where reliability and safety are top priorities, including diagnostic equipment, therapeutic equipment, and medical peripheral devices. Products Series Name:UMOYT3036 |
|
Products Details
| Size | L:101.6 W:50.8 H:29.6/mm |
|---|---|
| AC Input Rating | 90V~264V |
| DC Output Rating |
|
| Safety |
|
| Application Examples by Use Case | Introduction to Products by Application (PDF) |
If you are looking for a Products with specifications other than those listed above, please contact our sales team.
Contact Form
The UxOYT 36W Series (UxOYT3036) is an open-frame power supply (PCB power supply) for medical equipment, suitable for integration into devices. Designed for Medical/Home/I.C.T. applications, its space-saving 2" x 4" PCB size makes it easier to select an internal power supply based on mounting requirements.
This AC-DC switching power supply is designed to facilitate verification of power supply requirements for mass-production designs, including protection, EMC (emissions), and immunity. It features short-circuit, overvoltage, and overcurrent protection circuits and can be used to organize requirements based on Energy Efficiency Level VI (ErP/DoE), various EMC standards, and immunity criteria.
| Why Choose UNIFIVE: As a power supply manufacturer with more than 35 years of development experience, we provide comprehensive support, from specification review and certification to mass-production launch and post-production follow-up. Based on our standard products, we offer flexible supply to meet a range of volume requirements, from small-lot prototyping and evaluation to small- and medium-scale production and high-volume production.We also offer semi-custom solutions and support power supply selection from the specification review stage through mass-production adoption. |
Key Features
1) Open-Frame Power Supply (PCB Power Supply) for Medical Equipment
This open-frame power supply for medical equipment is designed for integration into devices as an internal power supply. Its space-saving 2" x 4" PCB size makes it easy to use even where mounting space inside the enclosure is limited.
2) Technical Standards and Protection Circuits
The applicable safety standards vary by model classification (M/H/N). Please check the details for each model. The product also complies with 100V standards. Protection circuits include short-circuit protection, overvoltage protection, and overcurrent protection, while overtemperature protection is optional.
3) Efficiency
The product supports Energy Efficiency Level VI (ErP/DoE), enabling selection based on the requirements of Commission Regulation (EU) 2019/1782 and DOE 10 CFR Parts 429 and 430.
4) EMC (Emissions)
The product supports FCC Part 15-B/FCC Part 18-B, EN (CISPR) 55032-B/55011-B/55014-B, BS EN 55032/55011/55014-1, and other standards. See "Standards and Compliance" for details.
5) Immunity
Based on the requirements of EN 55035/BS EN 55035, EN 60601-1-2, EN 55014-2, and other standards, immunity criteria under EN 61000-4-2/-3/-4/-5/-6/-8/-11 are specified. See "Standards and Compliance" for details.
Recommended Applications
- Integrating a 36W-class open-frame power supply (PCB power supply) for medical equipment as an internal power supply
- Selecting a power supply based on a 2" x 4" PCB size where mounting space is limited
- Reducing the risk of operational problems with a power supply equipped with short-circuit, overvoltage, and overcurrent protection
- Selecting a power supply in accordance with EMC emission requirements such as FCC, EN (CISPR), and BS EN
- Selecting a power supply based on immunity requirements (EN 61000-4-x), taking disturbances such as ESD, EFT, and surges into account
Application Examples
Suitable for a wide range of internal power supply applications, including medical equipment, home appliances, I.C.T. equipment, and AV equipment.
- Devices requiring an internal power supply
- Medical equipment
- Home appliances and consumer equipment
- Information and communications equipment/peripheral devices, etc.
* The above are examples and do not limit the possible applications.
Standards and Compliance
| Item | Details |
|---|---|
| Technical Standards | M: 60601-1/H: 61558-1, 60335-1/N: 62368-1, 60065-1, 60950-1; compliant with 100V standards |
| Efficiency | Energy Efficiency Level VI (ErP/DoE)/Commission Regulation (EU) 2019/1782/DOE 10 CFR Parts 429 and 430 |
| EMC (Emissions) | FCC Part 15-B/FCC Part 18-B/EN (CISPR) 55032-B/EN (CISPR) 55011-B/EN (CISPR) 55014-B/BS EN 55032/BS EN 55011/BS EN 55014-1 |
| Immunity | EN 55035/BS EN 55035/EN 60601-1-2/BS EN 60601-1-2/EN 55014-2/BS EN 55014-2, EN 61000-4-2/-3/-4/-5/-6/-8/-11 |
| Protection Circuits | Short-circuit protection/overvoltage protection/overcurrent protection/overtemperature protection (optional) |
| Operating Environment | Input (voltage): AC 100-240V (allowable range: AC 90-264V), input (frequency): 50/60Hz (allowable range: 47-63Hz)/operating temperature: 0-50°C, storage temperature: -20-65°C, operating/storage humidity: 5-95% RH (non-condensing) |
| Reference: Typical Models | 12.0V 3.0A (UxOYT3036-120030SA), 15.0V 2.4A (-150024SA), 19.0V 1.9A (-190019SA), 24.0V 1.5A (-240015SA), 48.0V 0.75A (-480750SA)/output accuracy: +/-5%, ripple/noise (examples: 240mV; 480mV for 48V) |
| Operating Environment (Input) | Input: AC 100-240V (allowable range: AC 90-264V), 50/60Hz (allowable range: 47-63Hz) |
| Reference: Dimensions and Connectors | Dimensions: 101.6 x 50.8mm, mounting holes: dia. 3.5, CN1: input (AC(L)/AC(N)), CN21: output (V-/V+) |
| Other | 2 x MOPP, LPS, RoHS 2 compliant, 5-year service life (optional) |