UMOYT3048 P Series Model supporting peak loads for 1 second
|
UMOYT3048 P is a 48 W class built-in open-frame power supply designed primarily for medical equipment applications and compliant with IEC 60601-1. Its 2 x MOPP-compliant insulation design ensures a high level of safety even in medical equipment intended for patient contact, enabling power supply designs that meet stringent medical safety requirements. In addition to its rated output, it features a peak power function capable of delivering up to 175% of the rated output for a specified duration, making it suitable for medical equipment applications involving motor drives, actuator startup, and inrush loads. Its high-efficiency design achieves efficiency of up to 87.77%, combining energy-saving performance with a stable power supply. It provides short-circuit, overvoltage, overcurrent, and overtemperature protection and complies with various EMC/immunity standards, including IEC 60601-1-2. Despite supporting peak power, its 2 x 4-inch board is designed for integration into medical equipment. It is a power supply solution that supports the design of diagnostic equipment, therapeutic equipment, medical peripherals, and other medical devices requiring both instantaneous high loads and a high level of safety. Products Series Name:UMOYT3048 P Series Model supporting peak loads for 1 second |
|
Products Details
| Size | L:101.6 W:50.8 H:30.6/mm |
|---|---|
| AC Input Rating | 90V~264V |
| DC Output Rating |
|
| Safety |
|
| Application Examples by Use Case | Introduction to Products by Application (PDF) |
If you are looking for a Products with specifications other than those listed above, please contact our sales team.
Contact Form
The UxOYT 48W P Series (UxOYT3048P) is an open-frame power supply (PCB power supply) for medical devices designed for Medical/Home/I.C.T. applications. Its space-saving 2" x 4" PCB design facilitates power supply selection based on the mounting requirements of embedded equipment.
This series is designed to accommodate peak loads (1 second), making it useful for defining power supply requirements for equipment with fluctuating loads. In addition to short-circuit, overvoltage, and overcurrent protection, overtemperature protection is also available as an option.
| Why Choose UNIFIVE: As a power supply manufacturer with more than 35 years of development experience, we provide comprehensive support from specification review and certification to mass-production launch and post-production follow-up. Based on our standard products, we offer flexible supply to meet a range of volume requirements, from small-lot prototyping and evaluation to small- and medium-scale production and high-volume production.We also offer semi-custom solutions and support power supply selection from the specification review stage through mass-production adoption. |
Key Features
1) Open-Frame Power Supply for Medical Devices (PCB Power Supply)
This open-frame power supply for medical devices is designed for integration as an embedded power supply. Its space-saving 2" x 4" PCB design makes it suitable for use even where installation space inside the enclosure is limited.
2) Technical Standards and Protection Circuits
The applicable safety standards differ by model category (M/H/N). Protection circuits include short-circuit, overvoltage, and overcurrent protection, while overtemperature protection is optional.
3) Supports Peak Loads (1 Second)
The peak-load function supports peak loads of up to 175% of the Rated Output (peak duration: 1 second, duty: 0.2). This facilitates requirement definition during design and evaluation while accounting for startup and load fluctuations.
4) Efficiency
Up to 85.00%.
5) EMC (Emissions)
Compliant with FCC Part15-B/Part18-B, EN(CISPR)55032-B/55011-B/55014-B, and BS EN 55032/55011/55014-1. See "Standards and Compliance" for details.
6) Immunity
Based on the requirements of EN55035/BS EN 55035, EN60601-1-2, EN55014-2, and other standards, immunity criteria are specified for EN61000-4-2/-3/-4/-5/-6/-8/-11. See "Standards and Compliance" for details.
Recommended Use Cases
- Integrating a 48W-class open-frame power supply for medical devices (PCB power supply) as an embedded power supply
- Selecting a power supply based on a 2" x 4" PCB size where installation space is limited
- Defining power supply requirements while accounting for peak loads (1 second) caused by startup, inrush loads, or other short-term load fluctuations
- Reducing operational risks with a power supply equipped with short-circuit, overvoltage, and overcurrent protection
- Selecting a power supply that meets emissions requirements such as FCC/EN(CISPR)/BS EN for EMC design
- Selecting a power supply based on immunity requirements (EN61000-4-x), taking ESD, EFT, surges, and other disturbances into account
Applications and Use Examples
Suitable for a wide range of embedded power supply applications, including medical devices, home appliances, I.C.T. equipment, and AV equipment.
- Equipment requiring an embedded power supply
- Medical devices
- Home appliances and consumer equipment
- Information and communications equipment/peripheral devices, etc.
* The above are examples and do not limit the possible applications.
Standards and Compliance
| Item | Details |
|---|---|
| Technical Standards | M: 60601-1/H: 61558-1 and 60335-1/N: 62368-1, 60065-1, and 60950-1; compliant with 100V standards (varies by model category) |
| Efficiency | Up to 85.00% |
| Peak Load | Up to 175% (peak duration: 1 second, duty: 0.2) |
| EMC (Emissions) | FCC Part15-B/FCC Part18-B/EN(CISPR)55032-B/EN(CISPR)55011-B/EN(CISPR)55014-B/BS EN 55032/BS EN55011/BS EN55014-1 |
| Immunity | EN55035/BS EN 55035/EN60601-1-2/BS EN60601-1-2/EN55014-2/BS EN55014-2, EN61000-4-2/-3/-4/-5/-6/-8/-11 |
| Protection Circuits | Short-circuit protection/overvoltage protection/overcurrent protection/overtemperature protection (optional) |
| Operating Environment | Input voltage: AC100-240V (allowable range: AC90-264V), input frequency: 50/60Hz (allowable range: 47-63Hz)/operating temperature: 0-50°C, storage temperature: -20-65°C, operating/storage humidity: 5-95%RH (non-condensing) |
| Reference: Representative Models | 12.0V 4.0A (peak: 7.0A), 15.0V 3.2A (peak: 5.6A), 19.0V 2.5A (peak: 4.37A), 24.0V 2.0A (peak: 3.5A), 48.0V 1.0A (peak: 1.75A)/output accuracy: +/-5%, ripple/noise: 240mV for 12-24V and 480mV for 48V |
| Operating Environment (Input) | Input: AC100-240V (allowable range: AC90-264V), 50/60Hz (allowable range: 47-63Hz) |
| Reference: Dimensions and Connectors | Dimensions: 101.6 x 50.8mm (tolerance: +/-0.5mm), maximum height: 26.0mm (MAX)/input CN1: equivalent to JST B3P-VH (AC(N)/AC(L))/output CN21: equivalent to JST B4P-VH (1-2: V-, 3-4: V+) |
| Other | 2 x MOPP, LPS, RoHS 2 compliant, 5-year service life (optional) |
Semi-Custom Solutions
Please feel free to contact us regarding specification confirmation, quotations, or samples. Examples of semi-custom solutions are available on the Customization Examples page.