UMOAT3060
|
The UMOAT3060 is a 60 W open-frame switching power supply for medical equipment. It complies with IEC 60601-1 and provides 4000 Vac dielectric strength, 2xMOPP patient protection, and 2xMOOP operator protection. Its double-fuse design, with fuses on both the L and N lines, eliminates the need for additional upstream components such as fuses and isolation Transformer, making it an ideal built-in power supply for medical equipment. Products Series Name:UMOAT3060 |
|
Products Details
| Size | L:105.0 W:60.0 H:31.6/mm |
|---|---|
| AC Input Rating | 90V~264V |
| DC Output Rating |
|
| Safety |
|
| Application Examples by Use Case | Introduction to Products by Application (PDF) |
If you are looking for a Products with specifications other than those listed above, please contact our sales team.
Contact Form
The UxOAT 60W Series (UxOAT3060) is an open-frame power supply for medical equipment (PCB power supply) designed for Medical/Home/I.C.T. applications. Intended for integration as an internal power supply, its space-saving 2.4" x 4.1" PCB design facilitates power supply selection based on mounting requirements.
As an AC-DC switching power supply, it features short-circuit, overvoltage, and overcurrent protection circuits and can be used for design and evaluation based on emission requirements (FCC/EN(CISPR)/BS EN) and immunity requirements (EN55035, EN61000-4-x, etc.).
| Why Choose UNIFIVE: As a power supply manufacturer with over 35 years of development experience, we provide comprehensive support from specification review and certification to mass-production launch and post-production follow-up. Based on our standard products, we offer flexible supply to meet a range of volume requirements, from small-lot prototyping and evaluation to small- and medium-scale production and high-volume production.We also offer semi-custom solutions and support power supply selection from the specification review stage through mass production. |
Key Features
1) Open-frame power supply for built-in medical equipment (PCB power supply)
This open-frame power supply for medical equipment is designed for integration as an internal power supply. Its space-saving 2.4" x 4.1" PCB design makes it suitable for use even where installation space inside the enclosure is limited.
2) Technical standards and protection circuits
The applicable safety standards vary by model category (M/H/N). The series also complies with the 100 V standard. Protection circuits include short-circuit, overvoltage, and overcurrent protection, with overtemperature protection also available as an option.
3) Efficiency
Compliant with Energy Efficiency Level VI (ErP/DoE), allowing selection based on the requirements of Commission Regulation (EU) 2019/1782 and DOE 10 CFR parts 429 and 430.
4) EMC (emissions)
Compliant with FCC (Part 15-B/Part 18-B), CE (EN(CISPR)55032-B/55011-B/55014-B), and BS EN (55032/55011/55014-1). See "Standards and Compliance" for details.
5) Immunity
Based on the requirements of EN55035/BS EN 55035, EN60601-1-2, EN55014-2, and other standards, immunity criteria are specified for EN61000-4-2/-3/-4/-5/-6/-8/-11. See "Standards and Compliance" for details.
Recommended Use Cases
- When integrating a 60 W-class open-frame power supply for medical equipment (PCB power supply) as an internal power supply
- When installation space is limited and power supply selection must be based on a 2.4" x 4.1" PCB size
- When reducing operational risks with a power supply equipped with short-circuit, overvoltage, and overcurrent protection
- When selecting a power supply in accordance with EMC emission requirements (FCC/EN(CISPR)/BS EN, etc.)
- When selecting a power supply based on immunity requirements (EN61000-4-x), taking external disturbances such as ESD, EFT, and surges into account
Applications and Examples
Suitable for a wide range of applications as an internal power supply, including medical equipment, home appliances, I.C.T. equipment, and AV equipment.
- Equipment requiring an internal power supply
- Medical equipment
- Home appliances and consumer electronics
- Information and communications equipment/peripheral devices, etc.
*The above are examples only and do not limit the possible applications.
Standards and Compliance
| Item | Details |
|---|---|
| Technical standards | M: 60601-1/H: 61558-1 and 60335-1/N: 62368-1, 60065-1, and 60950-1; compliant with the 100 V standard (varies by model category) |
| Efficiency | Energy Efficiency Level VI (ErP/DoE)/Commission Regulation (EU) 2019/1782/DOE 10 CFR parts 429 and 430 |
| EMC (emissions) | FCC Part 15-B/FCC Part 18-B/EN(CISPR)55032-B/EN(CISPR)55011-B/EN(CISPR)55014-B/BS EN 55032/BS EN55011/BS EN55014-1 |
| Immunity | EN55035/BS EN 55035/EN60601-1-2/BS EN60601-1-2/EN55014-2/BS EN55014-2, EN61000-4-2/-3/-4/-5/-6/-8/-11 |
| Protection circuits | Short-circuit protection/overvoltage protection/overcurrent protection/overtemperature protection (optional) |
| Operating environment | Input voltage: AC 100-240 V (allowable range: AC 90-264 V), input frequency: 50/60 Hz (allowable range: 47-63 Hz)/Operating temperature: 0-50°C, storage temperature: -20-65°C, operating/storage humidity: 5-95% RH (non-condensing) |
Semi-Custom Solutions
Please feel free to contact us to discuss specifications, quotations, or samples. Examples of our semi-custom solutions are available on the Customization Examples page.