UMOBT3060 Multi Output (2 Outputs)

This is a 60W open-frame switching power supply for medical equipment. Compliant with IEC60601-1, it supports 2xMOPP means of patient protection and 2xMOOP means of operator protection. It is a compact board-mounted power supply with dual outputs, making it an ideal model for embedded power applications in medical devices.

Product series name:UMOBT3060 Multi Output (2 Outputs)

This Products catalog (PDF)

General Products Catalog

UMOBT3060 Multi Output (2 Outputs)

Products details

Size L:127.0 W:76.0 H:30.6/mm
AC Input Rating 90V~264V
DC Output Rating
  • A:5.0V/2.5A&12.0V/4.0A
  • B:5.0V/2.5A&24.0V/2.0A
Safety
Examples of use by purpose Introduction of Products by Application (PDF)

If you are looking for products other than the above specifications, please contact our sales department.
Contact form

UxOBT 60W D Series (UxOBT3060D) is an open-frame power supply (PCB power supply) for medical equipment suitable for embedded applications. It supports dual output and features a series configuration (model classifications M / H / N) intended for Medical / Home / I.C.T. applications, making it easy to organize requirements as an internal power supply for equipment.

The PCB size is 3"×5", suitable for power selection based on mounting conditions. In addition to short-circuit, overvoltage, and overcurrent protection, overtemperature protection (optional) is also available. Emission and immunity items can be verified based on the standards listed in the specifications.

UNIFIVE Reasons to Choose Us:
As a power supply manufacturer with over 35 years of development experience, we provide comprehensive support from specification review with mass production in mind to certification acquisition, mass production launch, and post-production follow-up. Based on standard products, we offer flexible supply according to volume, from small-lot prototyping and evaluation to small- and medium-scale production, as well as large-scale mass production. We also support semi-custom solutions and assist with power supply selection from the specification review stage toward mass production adoption.

Key Features

1) Dual-output open-frame power supply (PCB power supply) for medical equipment

This dual-output open-frame power supply for medical equipment is designed for integration as an internal power supply. It can be utilized to organize power configuration requirements in equipment requiring multiple voltages.

2) Technical standards and protection circuits

Compliant with 62368-1 / 60601-1 / 61558-1 / 60335-1 and 100V standards (applicable standards vary by model classification). Equipped with short-circuit protection / overvoltage protection / overcurrent protection, with overtemperature protection available as an option.

3) Efficiency

Efficiency is specified as up to 80%.

4) EMC (Emissions)

Compliant with FCC Part15-B / Part18-B, EN(CISPR)55032-B / 55011-B / 55014-B, and BS EN 55032 / 55011 / 55014-1.

5) Immunity

Based on the requirements of EN55035 / BS EN 55035, EN60601-1-2, EN55014-2, etc., immunity items per EN61000-4-2 / -3 / -4 / -5 / -6 / -8 / -11 are specified.


Effective Use Cases


Applications

Widely applicable as an internal power supply for medical devices, home appliances, I.C.T., AV equipment, and more.

*The above are examples and do not limit applications.


Standards and Compliance

ItemDetails
Technical standards62368-1 / 60601-1 / 61558-1 / 60335-1, compliant with 100V standards (varies by model classification)
EfficiencyUp to 80%
EMC (Emissions)FCC Part15-B / FCC Part18-B / EN(CISPR)55032-B / EN(CISPR)55011-B / EN(CISPR)55014-B / BS EN 55032 / BS EN55011 / BS EN55014-1
ImmunityEN55035 / BS EN 55035 / EN60601-1-2 / BS EN60601-1-2 / EN55014-2 / BS EN55014-2, EN61000-4-2 / -3 / -4 / -5 / -6 / -8 / -11
Protection circuitsShort-circuit protection / overvoltage protection / overcurrent protection / overtemperature protection (optional)
Operating environmentInput (voltage): AC100-240V (tolerance: AC90-264V), Input (frequency): 50/60Hz (tolerance: 47-63Hz) / Operating temperature: 0-50°C, Storage temperature: -20-65°C, Operating/storage humidity: 5-95%RH (no condensation)
Reference: Representative models (dual output)5.0V 2.5A + 12.0V 4.0A, 5.0V 2.5A + 24.0V 2.0A (output accuracy ±5%, ripple 100mV, noise 150mV)
Reference: ConnectorsInput CN1: JSW A3961WV-5 (equivalent), Output CN2: JSW 6PIN A3961WV-6P (equivalent) - V1+ / V1-, V2+ / V2-
Reference: DimensionsPCB size 3"×5", External dimensions 127.0×76.0mm, Maximum height 26.0mm (MAX)
Other2×MOPP, LPS, RoHS 2 compliant, 5-year lifespan (optional)