UMOBT3060 Multi Output (2 Outputs)
|
This is a 60W open-frame switching power supply for medical equipment. Compliant with IEC60601-1, it supports 2xMOPP means of patient protection and 2xMOOP means of operator protection. It is a compact board-mounted power supply with dual outputs, making it an ideal model for embedded power applications in medical devices. Product series name:UMOBT3060 Multi Output (2 Outputs) |
|
Products details
| Size | L:127.0 W:76.0 H:30.6/mm |
|---|---|
| AC Input Rating | 90V~264V |
| DC Output Rating |
|
| Safety |
|
| Examples of use by purpose | Introduction of Products by Application (PDF) |
If you are looking for products other than the above specifications, please contact our sales department.
Contact form
UxOBT 60W D Series (UxOBT3060D) is an open-frame power supply (PCB power supply) for medical equipment suitable for embedded applications. It supports dual output and features a series configuration (model classifications M / H / N) intended for Medical / Home / I.C.T. applications, making it easy to organize requirements as an internal power supply for equipment.
The PCB size is 3"×5", suitable for power selection based on mounting conditions. In addition to short-circuit, overvoltage, and overcurrent protection, overtemperature protection (optional) is also available. Emission and immunity items can be verified based on the standards listed in the specifications.
| UNIFIVE Reasons to Choose Us: As a power supply manufacturer with over 35 years of development experience, we provide comprehensive support from specification review with mass production in mind to certification acquisition, mass production launch, and post-production follow-up. Based on standard products, we offer flexible supply according to volume, from small-lot prototyping and evaluation to small- and medium-scale production, as well as large-scale mass production. We also support semi-custom solutions and assist with power supply selection from the specification review stage toward mass production adoption. |
Key Features
1) Dual-output open-frame power supply (PCB power supply) for medical equipment
This dual-output open-frame power supply for medical equipment is designed for integration as an internal power supply. It can be utilized to organize power configuration requirements in equipment requiring multiple voltages.
2) Technical standards and protection circuits
Compliant with 62368-1 / 60601-1 / 61558-1 / 60335-1 and 100V standards (applicable standards vary by model classification). Equipped with short-circuit protection / overvoltage protection / overcurrent protection, with overtemperature protection available as an option.
3) Efficiency
Efficiency is specified as up to 80%.
4) EMC (Emissions)
Compliant with FCC Part15-B / Part18-B, EN(CISPR)55032-B / 55011-B / 55014-B, and BS EN 55032 / 55011 / 55014-1.
5) Immunity
Based on the requirements of EN55035 / BS EN 55035, EN60601-1-2, EN55014-2, etc., immunity items per EN61000-4-2 / -3 / -4 / -5 / -6 / -8 / -11 are specified.
Effective Use Cases
- When integrating a dual-output open-frame power supply for medical equipment as an internal power supply
- When organizing power configurations based on multiple voltages (e.g., 5V + 12V, 5V + 24V)
- When reducing operational risk with an internal power supply equipped with short-circuit / overvoltage / overcurrent protection
- When selecting a power supply in accordance with emission requirements (FCC / EN(CISPR) / BS EN) for EMC design
- When selecting a power supply considering immunity requirements (EN61000-4-x) against disturbances such as ESD, EFT, and surge
Applications
Widely applicable as an internal power supply for medical devices, home appliances, I.C.T., AV equipment, and more.
- Equipment requiring an internal power supply
- Medical devices
- Home and consumer electronics
- Information and communication equipment / peripheral devices, etc.
*The above are examples and do not limit applications.
Standards and Compliance
| Item | Details |
|---|---|
| Technical standards | 62368-1 / 60601-1 / 61558-1 / 60335-1, compliant with 100V standards (varies by model classification) |
| Efficiency | Up to 80% |
| EMC (Emissions) | FCC Part15-B / FCC Part18-B / EN(CISPR)55032-B / EN(CISPR)55011-B / EN(CISPR)55014-B / BS EN 55032 / BS EN55011 / BS EN55014-1 |
| Immunity | EN55035 / BS EN 55035 / EN60601-1-2 / BS EN60601-1-2 / EN55014-2 / BS EN55014-2, EN61000-4-2 / -3 / -4 / -5 / -6 / -8 / -11 |
| Protection circuits | Short-circuit protection / overvoltage protection / overcurrent protection / overtemperature protection (optional) |
| Operating environment | Input (voltage): AC100-240V (tolerance: AC90-264V), Input (frequency): 50/60Hz (tolerance: 47-63Hz) / Operating temperature: 0-50°C, Storage temperature: -20-65°C, Operating/storage humidity: 5-95%RH (no condensation) |
| Reference: Representative models (dual output) | 5.0V 2.5A + 12.0V 4.0A, 5.0V 2.5A + 24.0V 2.0A (output accuracy ±5%, ripple 100mV, noise 150mV) |
| Reference: Connectors | Input CN1: JSW A3961WV-5 (equivalent), Output CN2: JSW 6PIN A3961WV-6P (equivalent) - V1+ / V1-, V2+ / V2- |
| Reference: Dimensions | PCB size 3"×5", External dimensions 127.0×76.0mm, Maximum height 26.0mm (MAX) |
| Other | 2×MOPP, LPS, RoHS 2 compliant, 5-year lifespan (optional) |