UMDEC3120 UMI
|
This is an inlet-type 120W medical device AC/DC Power Adapter. Compliant with 2xMOPP means of patient protection and 2xMOOP means of operator protection. Meets U.S. efficiency standards, DOE Level VI, and is an ideal medical power supply for a wide range of medical devices. Product series name:UMDEC3120 UMI |
|
Products details
| Size | L:174.7 W:72.3 H:35.1/mm (IEC 320 C14) |
|---|---|
| AC Input Rating | 90V~264V |
| DC Output Rating |
|
| Safety |
|
| Examples of use by purpose | Introduction of Products by Application (PDF) |
If you are looking for products other than the above specifications, please contact our sales department.
Contact form
The UMDE 120W Series is a desktop-type AC/DC adapter (120W class) for medical equipment. It complies with the medical safety standard 60601-1 and supports 2xMOPP (Means of Patient Protection). The lineup includes one type, UMDEC3120, and as a desktop model with an AC inlet connection, it is suitable for stationary medical equipment applications.
Compliant with Energy Efficiency Level VI (ErP/DoE), as well as EU Commission Regulation (EU) 2019/1782 and DOE 10 CFR Part 429 and 430. This high-efficiency design achieves a power factor of 0.9 or higher. Equipped as standard with protection circuits for short circuit, overvoltage, and overcurrent (overtemperature protection available as an option). With an average active efficiency of 88.00% and no-load power consumption of 0.21W, four output voltage models ranging from 12.0V 10.0A to 24.0V 5.0A support a wide variety of medical devices.
| UNIFIVE Reasons to Choose Us: As a power supply manufacturer with over 35 years of development experience, we provide comprehensive support from specification review assuming mass production adoption, certification acquisition, production launch-up, to post-mass-production follow-up. Based on standard products, we offer flexible supply according to volume requirements, from small-lot prototyping and evaluation to small- and medium-scale production, as well as large-scale mass production. We also support semi-custom solutions and assist with power supply selection from the specification review stage toward mass production adoption. |
Key Features
1) Compliant with Medical Standard 60601-1 and Supports 2xMOPP
Compliant with the medical safety standard 60601-1 and also supports PSE Appendix 8. As a means of patient protection, it supports 2xMOPP (Means of Patient Protection). Designed to ensure high safety as an embedded power supply for medical equipment.
2) 120W Class High Output - Available in 4 Models
Available in four models: 12.0V 10.0A / 15.0V 8.0A / 19.0V 6.3A / 24.0V 5.0A. Delivering 120W-class high output, suitable for medical equipment requiring high power supply. Designed with high efficiency, achieving an average active efficiency of 88.00%.
3) High Power Factor Design (Power Factor 0.9 or Higher)
Achieves a power factor of 0.9 or higher at both AC100V and AC230V input. The superior power quality design contributes to efficient energy usage and reduced load on power infrastructure.
4) Energy Efficiency (ErP/DoE Level VI)
Energy Efficiency Level VI (ErP/DoE) compliant, meeting EU Commission Regulation (EU) 2019/1782 and DOE 10 CFR Part 429 and 430. High-efficiency design with no-load power consumption of 0.21W meets energy-saving requirements.
5) Protection Circuits and EMC Compliance
Equipped with short circuit, overvoltage, and overcurrent protection as standard, with overtemperature protection available as an option. For EMC standards, emissions comply with FCC Part18-B / EN(CISPR)55011-B / VCCI-B / BS EN55011, and immunity complies with EN60601-1-2 / BS EN60601-1-2 (EN61000-4-2 / 4-3 / 4-4 / 4-5 / 4-6 / 4-8 / 4-11).
Main Specifications
| Series Name | UMDE 120W Series |
| Products Category | Medical AC/DC Adapter - Desktop Type |
| Rated Output | 120W Class |
| Technical Standard (Safety) | 60601-1 |
| Patient Protection | 2xMOPP (Means of Patient Protection) |
| Energy Efficiency | Energy Efficiency Level VI (ErP/DoE), compliant with EU Commission Regulation (EU) 2019/1782 and DOE 10 CFR Part 429 and 430 |
| Emissions | FCC Part18-B / EN(CISPR)55011-B / VCCI-B / BS EN55011 |
| Immunity | EN60601-1-2 / BS EN60601-1-2 (EN61000-4-2 / 4-3 / 4-4 / 4-5 / 4-6 / 4-8 / 4-11) |
| Protection Circuits | Short circuit / Overvoltage / Overcurrent / Overtemperature (optional) |
| Operating Environment | Input: AC100~240V (range: AC90~264V), 50/60Hz (range: 47~63Hz), power factor 0.9 or higher (AC100V/AC230V) / Operating temperature: 0~40C (natural convection, above sea level) / Storage temperature: -20~65C (natural convection, above sea level) / Operating and storage humidity: 5~95%RH (no condensation) |
| Reference: Representative Models | 12.0V 10.0A / 15.0V 8.0A / 19.0V 6.3A / 24.0V 5.0A (±5%, ripple 240mV, noise 480mV, average active efficiency 88.00%, no-load power consumption 0.21W) |
| Reference: Type | UMDEC3120 (desktop type, AC inlet connection, 1 type) |
| Reference: Dimensions | 174.70x72.30x35.10mm |
| Other | RoHS 2 compliant |
Semi-Custom Support
Please feel free to contact us for specification confirmation, quotations, or sample requests. Examples of semi-custom solutions can be found on our Customization Case Studies page.